SD-BB-300(plus) Infant baby incubator with air and oxygen blender
It is the advanced technology in such disciplines as comprehensive clinical medicine,machinery, computer automatic control, sensor, etc. that the baby in incubator, offers newborn baby and disease baby to a fine environment where air purify, temperature and humidity is suitable, similar to uterus of the parent. The temperature of cabin space of the baby in incubator should be set up on the doctor's order, Baby's skin temperature, baby's air temperature of the cabin and baby's humidity of the cabin have digital display, when baby's skin temperature and air temperature of the cabin over
the normal value or present other unusual situations, there is alarming (power cut off, fan failure, sensor trouble, air
temperature over, skin temperature over, deviation between display temperature and setting up temperature, etc.), guarantee the dependability and security of the one machine.
This manual includes two parts of operation instruction and technical specification, and is suitable for BB-300Plus baby incubator. to install, use, maintain, common trouble analysis and get rid of, etc. product details Main technical parameter
Power Supply: ~ 220V 50/60Hz
Power consumption: 650VA (including power output socket: 200VA)
Operation environment conditions:
- An ambient Temperature: 20ºC~ 30ºC
- An ambient Humidity: 30% RH ~ 75% RH
- An ambient air pressure: 700 hPa ~1060 hPa
- An ambient air velocity less the 0.3m/s
Humidity display range: 10%~99%RH
Humidity control range: 40%~95%RH (humidity set value should higher 10% then ambient humidity)
Control temperature range:
- Air temperature : 25ºC~37ºC, 37.1ºC~38ºC
- Skin temperature : 32ºC~37ºC, 37.1ºC~38ºC
The warm-up time: ≤45 minutes
The function of alarm:
The over temp of air temp : 38ºC (Heater is cut-out simultaneously, the thermal cut-out should be reset by manual)
The second thermal cut-out temperature: 40ºC